A domestic medical equipment manufacturer has introduced the MC-100BY (model MCI0733), a 100mA portable X-ray diagnostic system designed to bring imaging closer to the patient's bedside. The unit is aimed at hospitals and clinics seeking greater flexibility in radiographic workflows and faster turnaround on routine diagnostic studies.
The system arrives as medical institutions place increasing emphasis on mobility and diagnostic efficiency in imaging departments, where fixed radiographic rooms can become a bottleneck for patients who cannot easily be moved.
The MC-100BY is intended for general radiographic acquisition, including imaging of the head, limbs and thoracic cavity. Its cantilever arm structure allows the unit to be maneuvered into wards and operating rooms, supporting bedside examinations where transporting a patient to a fixed imaging suite would be impractical.
Compared with conventional 50mA portable units, the 100mA power output offers better image quality and greater penetration, according to the manufacturer's specifications. That distinction matters in settings where patient anatomy or positioning limits the exposure settings available to the operator.
The device uses a single-focus, full-wave rectified X-ray generator with an integrated SCM control system, a configuration the manufacturer says reduces maintenance complexity. The control interface includes a high-resolution LCD screen that stores eight preset exposure parameter sets, which clinicians can recall, modify and save according to the anatomy being examined.
An SCR zero-point control circuit is also fitted to support stable and safe high-voltage output. The combination of stored protocols and a simplified control architecture is intended to shorten setup time between exposures.
| Technical parameter | Specification |
|---|---|
| Power supply requirements | 180–240 V, 50 Hz, ≥7 kVA |
| Exposure voltage | 50–100 kV |
| Exposure current | 16 mA, 32 mA, 63 mA, 100 mA |
| Exposure time | 0.08 s – 6.3 s |
| Focal spot size | 4.3 mm × 4.3 mm |
| Weight | Net 150 kg / gross 240 kg |
The design incorporates automatic exposure time control, filament preheating and component temperature monitoring. Together these functions are meant to protect the generator during repeated use and to keep exposure conditions within expected ranges.
Remote operation is supported at a maximum distance of seven meters, a feature that allows the operator to stand clear of the primary beam during exposure. The cantilever configuration also reduces the overall footprint of the machine, which the manufacturer cites as an advantage when moving through crowded corridors and confined clinical spaces.
Installations must supply between 180 and 240 V at 50 Hz. Because instantaneous current can reach 35 A, the power source requires an internal resistance below 1.0 ohm and a capacity of at least 7 kVA to maintain stability at high output.
At a net weight of 150 kg, the unit is transportable but not handheld, and facilities will need to confirm floor loading and doorway clearances along intended routes before deployment.